ORIGINAL ARTICLE  
Year : 2019  |  Volume : 44  |  Issue : 1  |  Page : 31-34
 

Comparison of performance of digital hemoglobinometer over automated hematology analyzer for hemoglobin estimation and its user-friendliness among the pregnant women in selected district hospitals of Madhya Pradesh


1 Department of Community Medicine, Gandhi Medical College, Bhopal, Madhya Pradesh, India
2 Department of Community Medicine, AIIMS, Raipur, Chhattisgarh, India
3 Maternal and Child Health, NHM, Government of MP, Madhya Pradesh, India

Date of Submission10-Jul-2018
Date of Acceptance02-Jan-2019
Date of Web Publication12-Mar-2019

Correspondence Address:
Dr. Devendra Gour
Department of Community Medicine, Gandhi Medical College, Bhopal, Madhya Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijcm.IJCM_216_18

Rights and Permissions

 

   Abstract 


Context: There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. Aims: The aim of this study was to compare two methods for hemoglobin estimation, i.e., automated hematology analyzer and Digital Hemoglobinometer, and to find out the sensitivity and specificity of Digital Hemoglobinometer for the estimation of hemoglobin. Subjects and Methods: A hospital-based cross-sectional study was carried out for 6 months from April to September 2017 in a District Hospital of five High Priority Districts of Madhya Pradesh. Two hundred and sixty antenatal females per district were selected for the study. Results: The mean hemoglobin by autoanalyzer is 10.19, and that by Digital Hemoglobinometer device is 9.89. Overall, sensitivity of Digital Hemoglobinometer for hemoglobin estimation was calculated to be 89.4% and specificity was calculated to be 63.6%. Positive predictive value was found to be 82.6% and negative predictive value was 75.8% compared against AutoAnalyser (gold standard). Conclusions: As the Digital Hemoglobinometer device has high sensitivity and specificity and good diagnostic accuracy, it must be used at the community level in resource-poor setting for hemoglobin estimation. In primary health-care conditions, Digital Hemoglobinometer can significantly reduce misdiagnosis of anemia compared with clinical assessment alone.


Keywords: Anemia, antenatal clinic female, Auto Analyser, Digital Hemoglobinometer, Madhya Pradesh


How to cite this article:
Toppo M, Pal DK, Gour D, Melwani V, Dubey M, Mishra A. Comparison of performance of digital hemoglobinometer over automated hematology analyzer for hemoglobin estimation and its user-friendliness among the pregnant women in selected district hospitals of Madhya Pradesh. Indian J Community Med 2019;44:31-4

How to cite this URL:
Toppo M, Pal DK, Gour D, Melwani V, Dubey M, Mishra A. Comparison of performance of digital hemoglobinometer over automated hematology analyzer for hemoglobin estimation and its user-friendliness among the pregnant women in selected district hospitals of Madhya Pradesh. Indian J Community Med [serial online] 2019 [cited 2019 Mar 26];44:31-4. Available from: http://www.ijcm.org.in/text.asp?2019/44/1/31/253906





   Introduction Top


Assessment of hemoglobin is one of the most reliable indicators for anemia and is widely used to screen for anemic individuals and to evaluate responses to interventions.[1] Commonly used methods to estimate hemoglobin in a community setting are clinical examination for pallor, Sahli's method, World Health Organization color scale, and HemoCue. Unfortunately, these methods have several limitations, ranging from the lack of accuracy to complexity to high cost.[2],[3] There are various methods of hemoglobin estimation, invasive and noninvasive, of which the invasive type varies from simple paper scale reading to measurement by photometer, i.e., HemoCue and Sahli's method. The noninvasive types are the pulse oximetry, photoplethysmography, optoacoustic method, diffuse reflectance spectroscopy, and imaging-based technique, each with its own advantages and limitations. Accurate quantitative point of care diagnostic tests can confirm the diagnosis of anemia through measurement of a decreased amount of red blood cells or decreased hemoglobin concentration in the blood, but these are not suitable in most primary health-care settings with very low resources because they either require constant quality control by trained staff, use toxic or expensive reagents and consumables, or depend on electricity supply.[4] Diagnosis is thus often based on clinical signs alone such as conjunctival, palmar, and nail bed pallor. None of these signs, whether combined or singly, yield an acceptable diagnostic accuracy.[5] This leaves many cases undetected and untreated and also possesses the risk of unnecessary and potentially harmful blood transfusions, increasing the risk of transmission of blood-borne pathogens, and wasting resources in case of misdiagnosed severe anemia.[6]

There is a need for a simple screening method for the detection of anemia that can be used by public health workers in the field. Any method of screening or monitoring individuals for anemia at primary care level should be cheap, simple to operate, sturdy enough for field use, dependent neither on electricity nor batteries, and reasonably accurate. It should also use a minimum of materials that require regular replacement and should give immediate results. Hemoglobin concentration is routinely measured using Automated Hematology Analyzers (AHAs). Although these are very accurate and reliable, they are expensive, and problems of samples' transport to the laboratory may delay treatment.[7] In clinical measurement, comparison of a new measurement technique with an established one is often needed to see whether they agree sufficiently for the new to replace the old. Therefore, this study was aimed to compare two hemoglobin testing methods and to assess the utility of Digital Hemoglobinometer against a standard hematology AutoAnalyzer and to ascertain whether Digital Hemoglobinometer method could replace the traditional hematology AutoAnalyzer for hemoglobin screening. The objectives of the study were (1) to compare two methods for hemoglobin estimation, i.e., AHA and Digital Hemoglobinometer and (2) to find out the sensitivity and specificity of Digital Hemoglobinometer for the estimation of hemoglobin.


   Subjects and Methods Top


It was a hospital-based cross-sectional study carried out for 6 months from April to September 2017 in a District Hospital of five High Priority Districts of Madhya Pradesh were selected based on case load and availability of AutoAnalyzer. The study participants were antenatal mothers attending the antenatal clinics of selected district hospital. The sample size was estimated according to the prevalence of anemia, that is, 54.6% (NFHS IV MP). Assuming a sensitivity of 80% and specificity of 80% and precision of 0.03, a sample size of 1265 was considered adequate for the study. Thus, it was estimated to be 253 antenatal females per district which rounded off to 260 tests (antenatal mothers) was covered to assess the diagnostic accuracy of each method. Thus, a total of 1300 antenatal females were included in the study. Inclusion criteria were (1) pregnant women attending the antenatal clinic for the first time in the study period and (2) pregnant women giving consent for the study. Pregnant women attending the hospital with severe illness were excluded from the study. Sample collection – blood samples were collected from consented participants within the study group reporting to the hospital after study procedures had been explained to them. Following aseptic precautions, the venous blood was drawn in the ethylenediaminetetraacetic acid (EDTA) vial using disposable syringes. Venous blood collected from the participants was processed to estimate their Hb content using two different methods. Two investigators were positioned in each district at the site of sample collection to ensure blinding. Time-to-time calibration of the machine was ensured. Ethical approval was obtained from the Institutional Ethical Committee for the study. Statistical analysis – The data were entered into the Microsoft Excel 2007, and the analysis was done using Epi Info™ version 7.2.2.6 software (CDC). Epi Info™ is a trademark of the Centers for Disease Control and Prevention (CDC). The software is in the public domain and freely available for use, copying translation and distribution.[8]

AutoAnalyzer – The AutoAnalyzer is an automated blood cell counter intended for in vitro diagnostic use in clinical laboratories. It measures the hemoglobin concentration using a noncyanide hemoglobin method. The instrument has been proven to provide accurate and reliable results including hemoglobin concentrations. The test is performed by collecting 2 ml of blood in an EDTA vial using disposable syringe under all aseptic precautions. The test is performed as stated in the manufacturer's manual using the reagents/kits provided with the instrument as recommended by manufacturers. The AutoAnalyzer used in the study was manufactured by Celltac by Nihon Kohden MEK 6420P (Sagar), Mindray BC 2800 (Raisin), Mindray BC 5000 (Shahdol), Mindray BC 5300 (Anuppur), and URIT 2900 (Damoh).[9]

Digital Hemoglobinometer – It is a device in response to the need for a “simple, cheap, and robust device to measure hemoglobin by health workers outside the laboratory.” Digital Hemoglobinometer (HCG TRIESTA laboratory) is palm-sized nanobioelectronic device with self-calibration sensors that takes <60 s for each hemoglobin estimation. The Digital Hemoglobinometer System is based on the principle of reflectance photometry. Capillary, venous, or arterial whole-blood sample can be used for the hemoglobin estimation with the requirement of only 8 ml of blood sample. It has a rechargeable battery of 3.6 V that makes its suitable for usage in places with poor electricity supply. This device can be used in temperature range of 5–45°C. The device can store up to 1000 results for date and time, and easy record maintenance is possible using mobile application. The range of measurement is 0–25 mg/dl.[10] The Digital Hemoglobinometer Strips are thin plastic strips which contain chemical reagents. The strips vial has a unique code which needs to be entered in the device each time the strip is loaded. The pack also contains 50 sterile lancets.


   Results Top


The women attending the antenatal clinic were mostly of the age group of 24–26 years (32.7%) followed by 21–23 years (31.5%), <20 years (17.1%), and 27–29 years (10.8%). Only 7.9% of women belonged to the age group of 30 years and above.

The mean hemoglobin by AutoAnalyzer is 10.19, and that by Digital Hemoglobinometer device is 9.89. Median by AutoAnalyzer is 12.4 and by Digital Hemoglobinometer device is 13.65. There is only slight difference in the minimum values found by them. Z-score of two means of both the methods, i.e., AutoAnalyzer and Digital Hemoglobinometer, was calculated, and the result came out to be statistically nonsignificant. Hence, we can say that the observed difference between the two methods is just by chance.

[Table 1] shows the number of women diagnosed under each category by the two methods. The tables show that 34.07% of the women were classified as no anemia by AutoAnalyzer, whereas 28.61% were not anemic by Digital Hemoglobinometer. Furthermore, 25.61%, 37.15, and 3.15% were classified as mild, moderate, and severe anemia, respectively, with AutoAnalyzer, whereas 22.53%, 43.53%, and 5.30% women were classified as mild, moderate, and severe anemia, respectively, by Digital Hemoglobinometer. The reference test, i.e., AHA detected more number of no anemia and mild anemia, whereas the index test, i.e., Digital Hemoglobinometer, detected more cases of moderate and severe anemia.
Table 1: Distribution according to the severity of anemia as per the World Health Organization Class

Click here to view


Sensitivity, specificity, and positive and negative predictive values (NPVs) of Digital Hemoglobinometer were calculated over AutoAnalyzer (gold standard). [Table 2] shows that overall sensitivity of Digital Hemoglobinometer was calculated to be 89.4% and specificity was calculated to be 63.6%. Positive predictive value was found to be 82.6% and NPV was 75.8% compared against AutoAnalyzer (gold standard). For detection of mild anemia, sensitivity was calculated to be 60.7% and specificity was 72.4%. This means that the Digital Hemoglobinometer device is 60.7% sensitive to detect mild anemia. Similarly, Digital Hemoglobinometer is 97.9% sensitive to detect moderate anemia. Furthermore, the sensitivity of Digital Hemoglobinometer is 100%, and specificity is 98% for the detection of severe anemia, thereby indicating that Digital Hemoglobinometer is a very effective device in identifying anemia.
Table 2: Sensitivity and specificity of Digital Hemoglobinometer as compared to an AutoAnalyzer

Click here to view



   Discussion Top


Accurate determination of hemoglobin concentration is a common element in assessing the extent of anemia and making a decision regarding treatment. This decision should be made based on the reliable and rapidly assessed laboratory tests. The Digital Hemoglobinometer is a portable device for measuring hemoglobin concentration, and it requires very little staff training thus making it a very useful tool in resource-limited areas such as field conditions since it can easily be transported. In this study, we compared it with AutoAnalyzer used in the laboratory. We found no significant differences in the hemoglobin concentrations determined by the two methods. Our study is in agreement with other studies conducted in other settings to support the use of the device. These include the studies conducted by von Schenck et al. which suggested that HemoCue equipment is easy to handle.[11] Van de Louw et al., among patients with gastrointestinal bleeding, found that the mean difference between HemoCue and hemoglobin (bias) was −0.06 g/dl and standard deviation (precision) was 0.87 g/dl (95% Confidence interval – 1.8 − 1.68).[12] Rippmann et al., among surgical patients, in their study concluded that HemoCue underestimates the Hb concentration as determined by a CO-Oximeter by 2-5% and exhibits a significantly higher variability.[13] Neville RG et al., within urban general practice, found that the mean hemoglobin concentration by HemoCue was found to be 137 ± 23 g/l (range 64–192). The corresponding laboratory figures were mean 135 ± 1 and range 78–180.[14] Rechner et al., among neonates;[15] Lardi et al., among patients undergoing aortic surgery in the theater;[16] Sari et al., among Indonesian mothers;[17] and Radtke et al., among blood donors,[18] also showed the utility of Digital Hemoglobinometer comparable to gold standard. Whereas, studies conducted by Zhou et al. among pregnant women in a higher altitude area of Tibet, China,[19] and Bhaskaram et al., among apparently healthy children of 1–6 years[20] do not support the use of the HemoCue in their various study populations. Since Digital Hemoglobinometer is portable, simple to use, and gives results immediately, it is recommended as an excellent screening method for the detection of anemia in cellular Hb concentration, primary health care (PHC), blood bank, and field setup. However, Digital Hemoglobinometer costs a considerable amount of money, and this should be considered while planning and budgeting for data collection.


   Conclusions Top


As the Digital Hemoglobinometer device has high sensitivity and specificity and good diagnostic accuracy, it must be used at the community level in resource-poor setting. In PHC conditions, Digital Hemoglobinometer can significantly reduce misdiagnosis of anemia compared with clinical assessment alone.

Acknowledgment

I extend my sincere thanks to Mission Director NHM, Dr. Archana Mishra, and Dr. Shalini Kapoor, for granting the study. I acknowledge CMHO, MCH consultants, Lab Technicians for their excellent support and co-operation for providing valuable information of respective districts. Finally, I extend my special thanks to all antenatal mothers, ANM Mentors, and ASHA, without whom the study would not have been completed.

Financial support and sponsorship

This study was financially supported by NHM MP.

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Morris SS, Ruel MT, Cohen RJ, Dewey KG, de la Brière B, Hassan MN, et al. Precision, accuracy, and reliability of hemoglobin assessment with use of capillary blood. Am J Clin Nutr 1999;69:1243-8.  Back to cited text no. 1
    
2.
World Health Organization. Assessing the Iron Status of Populations. Geneva, Switzerland: World Health Organization; 2007.  Back to cited text no. 2
    
3.
Nestel P, Taylor H. Anemia Detection Methods in Low-Resource Settings: A Manual for Health Workers. Seattle, Washington, USA; Program for Appropriate Technology in Health; 1997.  Back to cited text no. 3
    
4.
Briggs C, Kimber S, Green L. Where are we at with point-of-care testing in haematology? Br J Haematol 2012;158:679-90.  Back to cited text no. 4
    
5.
Chalco JP, Huicho L, Alamo C, Carreazo NY, Bada CA. Accuracy of clinical pallor in the diagnosis of anaemia in children: A meta-analysis. BMC Pediatr 2005;5:46.  Back to cited text no. 5
    
6.
Marn H, Critchley JA. Accuracy of the WHO haemoglobin colour scale for the diagnosis of anaemia in primary health care settings in low-income countries: A systematic review and meta-analysis. Lancet Glob Health 2016;4:e251-65.  Back to cited text no. 6
    
7.
Jahr JS, Lurie F, Driessen B, Davis JA, Gosselin R, Gunther RA, et al. The HemoCue, a point of care B-hemoglobin photometer, measures hemoglobin concentrations accurately when mixed in vitro with canine plasma and three hemoglobin-based oxygen carriers (HBOC). Can J Anaesth 2002;49:243-8.  Back to cited text no. 7
    
8.
9.
10.
von Schenck H, Falkensson M, Lundberg B. Evaluation of “HemoCue,” a new device for determining hemoglobin. Clin Chem 1986;32:526-9.  Back to cited text no. 10
    
11.
Van de Louw A, Lasserre N, Drouhin F, Thierry S, Lecuyer L, Caen D, et al. Reliability of HemoCue in patients with gastrointestinal bleeding. Intensive Care Med 2007;33:355-8.  Back to cited text no. 11
    
12.
Rippmann CE, Nett PC, Popovic D, Seifert B, Pasch T, Spahn DR, et al. Hemocue, an accurate bedside method of hemoglobin measurement? J Clin Monit 1997;13:373-7.  Back to cited text no. 12
    
13.
Neville RG. Evaluation of portable haemoglobinometer in general practice. Br Med J (Clin Res Ed) 1987;294:1263-5.  Back to cited text no. 13
    
14.
Rechner IJ, Twigg A, Davies AF, Imong S. Evaluation of the HemoCue compared with the coulter STKS for measurement of neonatal haemoglobin. Arch Dis Child Fetal Neonatal Ed 2002;86:F188-9.  Back to cited text no. 14
    
15.
Lardi AM, Hirst C, Mortimer AJ, McCollum CN. Evaluation of the HemoCue for measuring intra-operative haemoglobin concentrations: A comparison with the Coulter Max-M. Anaesthesia 1998;53:349-52.  Back to cited text no. 15
    
16.
Sari M, de Pee S, Martini E, Herman S, Sugiatmi H, Bloem MW, et al. Estimating the prevalence of anaemia: A comparison of three methods. Bull World Health Organ 2001;79:506-11.  Back to cited text no. 16
    
17.
Radtke H, Polat G, Kalus U, Salama A, Kiesewetter H. Hemoglobin screening in prospective blood donors: Comparison of different blood samples and different quantitative methods. Transfus Apher Sci 2005;33:31-5.  Back to cited text no. 17
    
18.
Zhou X, Yan H, Xing Y, Dang S, Zhuoma B, Wang D, et al. Evaluation of a portable hemoglobin photometer in pregnant women in a high altitude area: A pilot study. BMC Public Health 2009;9:228.  Back to cited text no. 18
    
19.
Bhaskaram P, Balakrishna N, Radhakrishna KV, Krishnaswamy K. Validation of hemoglobin estimation using Hemocue. Indian J Pediatr 2003;70:25-8.  Back to cited text no. 19
    
20.
Centers for Disease Control and Prevention. “Epi info 7.” Atlanta: CDCP. 2011.  Back to cited text no. 20
    



 
 
    Tables

  [Table 1], [Table 2]



 

Top
Print this article  Email this article
           

    

 
   Search
 
  
    Similar in PUBMED
    Search Pubmed for
    Search in Google Scholar for
  Related articles
    Article in PDF (307 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
   Subjects and Methods
   Results
   Discussion
   Conclusions