LETTER TO EDITOR Table of Contents   
Year : 2009  |  Volume : 34  |  Issue : 4  |  Page : 367-368
 

Intra-rater and inter-rater reliability of hemoglobin color scale method


1 Department of Community Medicine, Faculty of Medicine and Allied Sciences, Rajarata University of Sri Lanka, Sri Lanka
2 Department of Community Dentistry, Faculty of Dental Science, University of Peradeniya, Peradeniya, Sri Lanka
3 Department of Community Medicine, Faculty of Medical Sciences, University of Sri Jayewardenepura, Nugegoda, Sri Lanka

Date of Submission19-Feb-2008
Date of Acceptance03-Jun-2009
Date of Web Publication17-Dec-2009

Correspondence Address:
S B Agampodi
Department of Community Medicine, Faculty of Medicine and Allied Sciences, Rajarata University of Sri Lanka
Sri Lanka
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-0218.58405

Rights and Permissions

 



How to cite this article:
Agampodi S B, Kularathna M, Pathiraja P. Intra-rater and inter-rater reliability of hemoglobin color scale method. Indian J Community Med 2009;34:367-8

How to cite this URL:
Agampodi S B, Kularathna M, Pathiraja P. Intra-rater and inter-rater reliability of hemoglobin color scale method. Indian J Community Med [serial online] 2009 [cited 2019 Sep 17];34:367-8. Available from: http://www.ijcm.org.in/text.asp?2009/34/4/367/58405


Sir,

Anemia during pregnancy is a major public health problem in Sri Lanka as well as in other developing countries. The clinical method of detecting anemia is having a low accuracy. [1],[2] The National Maternal and Child Health Program - Sri Lanka, has introduced the Hemoglobin Color Scale (HCS) to assess Hb levels in field clinics since 2007. The HCS test has been developed by the WHO, as it is a simple, reliable, and inexpensive method of diagnosing anemia when laboratory methods are not freely available. [3] The objective of the present study is to test the intra- and inter-rater reliability of HCS tests in the actual field setting.

We conducted the present study in Kalutara district, Sri Lanka. Five randomly selected pregnant mothers were invited for the study. Five public health midwives (PHMM) from the field were selected randomly and invited for the study. We used a balanced randomized block design, with repeated observations, for reliability testing. A single mother was tested using five strips. Reading of Hb values were done by PHMM who were blinded for the identity of strips. Five Hb strips representing all five subjects were given to each PHMM for recording of HB values. Strip sets were exchanged and reread by different PHMMs. The same two sets they already observed were given to them for a repeated reading without their knowledge.

One hundred readings were obtained in all. A single PHMM recorded four readings [two repetitive readings] for a subject, thus 20 recordings for all five subjects. Readings were completed within five minutes. Percentage agreement between two readings for an individual PHMM varied from 20 to 90%. [Table 1] shows the various measures of intra-rater reliability.

Variance of the repeated reading shows wide variation, that is, 0789 to 2.221 between PHMMs. Only two observers had very good Intraclass Correlation Coefficient (ICC) (0.757), which was highly significant (P = 0.004). One observer reported non-significant poor correlation (ICC 0.301, P = 0.184), while two remaining observers reported no correlation at all between repeated readings.

A single strip had four readings. Descriptive analysis of strips yielded that only 4/25 strips (16%) had 100% concordance, while 14 (56%) had 75% concordance rate.

ICC for test / retest reliability for strips was 0.162, where the correlation was very poor and not significant (P = 0.699). Repeated measure analysis of variance (ANOVA) showed highly significant differences between observers (F = 10.05, P < 0.001) and negative ICC value (- 0.047), which indicated that there was no correlation at all between the observers.

Our study showed that the inter-rater and intra-rater reliability was very poor for HCS, among our study sample. However, previous studies have shown that the HCS test has high reliability and validity. [4] The present study was conducted in a natural field setting, during their routing procedures, to resemble the actual environment where they usually apply this test. Following of standard procedure was not observed because our prime objective was to assess the present practice. These limitations might have an impact on our observations.

Retraining of PHMM in a standard procedure and identification of PHMM with reliable strip reading capabilities and allocation of the task only to those, would help to overcome this problem. However, the study was a preliminary study. Generalization of the findings cannot be done due to limitation in the study. A study with properly calculated sample size, with an adequate number of pregnant mothers representing a wide range of HB levels is recommended.

 
   References Top

1.Luby SP, Kazembe PN, Redd SC, Ziba C, Nwanyanwu OC, Hightower AW, et al. Using clinical signs to diagnose anaemia in African children. Bull World Health Organ 1995;73:477-82.  Back to cited text no. 1      
2.Sanchez-Carrillo CI. Bias due to conjunctiva hue and the clinical assessment of anaemia. J Clin Epidemiol 1989;42:751-4.  Back to cited text no. 2  [PUBMED]  [FULLTEXT]  
3.Stott GJ, Lewis SM. A simple and reliable method for estimating haemoglobin. Bull World Health Organ 1995;73:369-73.  Back to cited text no. 3  [PUBMED]  [FULLTEXT]  
4.Gosling R, Walraven G, Manneh F, Bailey R, Lewis SM. Training health workers to assess anaemia with the WHO haemoglobin colour scale. Trop Med Int Health 2000;5:214-21.  Back to cited text no. 4  [PUBMED]  [FULLTEXT]  



 
 
    Tables

  [Table 1]



 

Top
Print this article  Email this article
                 

    

 
   Search
 
  
    Similar in PUBMED
    Search Pubmed for
    Search in Google Scholar for
    Article in PDF (339 KB)
    Citation Manager
    Access Statistics
    Reader Comments
    Email Alert *
    Add to My List *
* Registration required (free)  


    References
    Article Tables

 Article Access Statistics
    Viewed1156    
    Printed91    
    Emailed0    
    PDF Downloaded104    
    Comments [Add]    

Recommend this journal

  Sitemap | What's New | Feedback | Copyright and Disclaimer
  2007 - Indian Journal of Community Medicine | Published by Wolters Kluwer - Medknow
  Online since 15th September, 2007